OCV test machine stand-alone equipment configuration table |
|||||
NO |
Main part name |
Model Specifications |
brand |
quantity |
Remark |
1 |
OCV Tester |
Japanese home 3561 |
Hioki |
1 set |
|
2 |
Scanning mechanism |
|
Honeywell |
1 set |
|
3 |
OCV testing agency |
|
Herolaser |
1 set |
|
4 |
Linear module + motor |
|
HIWIN |
3 sets |
|
5 |
Good product temporary storage tray |
|
Herolaser |
1 set |
|
6 |
NG product temporary storage tray |
|
Herolaser |
1 set |
|
7 |
upper and lower racks |
|
Herolaser |
1 set |
|
8 |
electrical control system |
Including PLC control program and some electrical components |
Siemens/Mitsubishi/Omron |
1 set |
OCV testing machine single machine energy consumption table |
|||||
Classification |
total machine power |
compressed air |
vacuum gas |
Protective gas |
pure water |
data |
3.0KW |
2L/min |
/ |
/ |
/ |
Main parameters of OCV tester:
Impedance measurement range | 30mΩ (maximum display: 31.000mΩ, resolution: 1μΩ, test current: 100mA)
300mΩ (maximum display: 310.00mΩ, resolution: 10μΩ, test current: 10mA) 3Ω (maximum display: 3.1000Ω, resolution: 100μΩ, test Current: 1mA) 30Ω (maximum display: 31.000Ω, resolution: 1mΩ, test current: 100μA) 300Ω (maximum display: 310.00Ω, resolution: 10mΩ, test current: 10μA) 3kΩ (maximum display: 3.1000kΩ, resolution: 100mΩ, test current: 10μA)
Measurement representative accuracy: ±0.5%rdg ±5 dgt Test current frequency: 1 kHz ± 0.2 Hz Measurement method: AC 4-terminal method |
Voltage measurement range | 6V (maximum display: 6.00000V, resolution: 10μV)
60V (maximum display: 60.0000V, Resolution: 100μV) Measurement representative accuracy: ±0.01%rdg ±3 dgt |
Measure response time | 10ms |
sampling time | Ω or V (60 Hz): 4 ms (EX.FAST), 12 ms (FAST), 35 ms (MEDIUM), 150 ms (SLOW)
Ω V (60 Hz): 8 ms (EX.FAST), 24 ms (FAST), 70 ms (MEDIUM), 253 ms (SLOW) Ω or V (50 Hz): 4 ms (EX.FAST), 12 ms (FAST), 42 ms (MEDIUM), 157 ms (SLOW) Ω V (50 Hz): 8 ms (EX.FAST), 24 ms (FAST), 84 ms (MEDIUM), 259 ms (SLOW) |
Function | Contact check function, zero adjustment function, pulse measurement function, comparator function (Hi/IN/Lo), statistical calculation function (maximum 30,000 sets of data), delay function, average function, setting save function (panel save), storage function LabVIEW ® driver correspondence |
interface | LAN (TCP/IP, 10BASE-T/100BASE-TX)
RS-232C (Max. 38.4 kbps, used for both printer I/F) EXT I/O (37-pin Handler interface) Analog output (DC 0V~3.1V) |
power supply | AC 100 V ~ AC 240 V, 50 Hz/60 Hz, 35 VA max. |
size and weight | 215W * 80H * 295D mm, 2.4 kg |